Cleaning the Reagent Probe Drains on the Dimension® Family of Integrated Chemistry Systems Video
This video provides instructions on how to clean probe drains on the Dimension® Family of Integrated Chemistry Systems.
From the operating menu with the instrument in standby, select the pause key. Next, raise the IMT and reagent lids. Turn the bottom gold colored spline shaft on the R2 reagent arm to raise the reagent probe out of the reagent drain. Move the R2 arm away from the drain by moving it counterclockwise. Turn the bottom gold colored spline shaft on the R1 reagent arm to raise the R1 reagent probe out of the reagent drain. If applicable to your system, turn the bottom gold colored spline shaft on the R3 reagent arm to raise the R3 reagent probe out of the reagent drain. Turn off the vacuum system before proceeding. From the operating menu, select. F7 diagnostics F1 electromechanical F6 waterworks, then select F4 vacuum on. Slash off twice to turn the vacuum off. Place a paper towel under the probes to catch any overflow using hemostats. Clamp off the nonlabeled drain line directly at the R1R2 and R3 drains if installed. Place a paper towel near the drain to catch any overflow. The R2 drain is shown here using a disposable plastic pipette filled with undiluted bleach or sample probe cleaner. Insert the pipette tip almost to the bottom of the drain and dispense the bleach into the drain unit until it is filled with liquid. Notice there are two ports here. This is also the case for the R3 drain. Remove the pipette and let the bleach sit in the drain for 5 minutes. Complete this step for each reagent probe drain. Take care not to overflow the R1 drain as it can destroy the photometer home sensor. Then unclap the drain line. Remove any paper towels. And close the instrument lids. Select F4 Vacuum on slash off to turn the system vacuum back on. From the operating menu, select F4 System Prep, then select F7 Pump Prime to access the pump prime screen. Change the cycle count to 10 and select Enter. Select F1 Prime Water. Be sure to change the cycle count back to the original number when priming is complete. If documenting using the paper logs from the Dimension System Operator's Guide, document the maintenance in the weekly monthly maintenance log sheet or instrument log sheet. Or, if documenting using the electronic logs in the Dimension system software, document the maintenance in the instrument log screen.
05:00 04:59 04:58 04:57 04:56 Cleaning the Reagent Probe Drains on the Dimension® Family of Integrated Chemistry Systems Video SIEMENS Standby Sampler Idle IMT OK 09/14/22 21:23 Dimension Exit Enter labmanager en OPERATING MENU STAT Status PLEASE SELECT AN OPTION USING THE FUNCTION KEYS: Sample Alert At any time, press HELP to get help, or press EXIT to exit a program. QC Alert Calib Alert To Shut down or Change Operator, press Sign Out or Alt+EXIT. Run Home F1: ENTER DATA F2: SAMPLE STATUS F3: TEST RESULTS F4: SYSTEM PREP F5: PROCESS CTRL F6: SYSTEM CONFIG F7: DIAGNOSTICS F8: SEGMENT STATUS F2 Lock CA Dimension EXL 20 Healthineers Dimension EXL 200 WP2 TO Note: This step is only applicable for the Dimension® RXL Max® Integrated Chemistry Systems. AUTION RMS (R3) Probe 2-PHASE DC 4.5V 2A Note: For the Dimension EXL with LM System, open the Reagent Management System lid to access R3 reagent arm. SE XP7 22854 06/25/23 18:12 DIAGNOSTIC MENU Press EXIT to exit a program. F1: ELECTRO/MECH F2: FILM LOAD F3: ALIGNMENTS F4: F5: CHECK WINDOWS F6: ERROR LOG F7: KINETICS F8: OPEN CHANNELS Diagnostics ELECTRO/MECHANICAL DIAGNOSTICS SELECT PROGRAM USING FUNCTION KEYS F1: REAGENT F2: SAMPLE SYSTEM F3: IMT F4: CUVETTE F5: PHOTOMETER F6: WATER WORKS F7: POWER SYSTEM F8: ULTRASONICS WATER & WASTE DIAGNOSTIC SENSORS: Water Empty: ON Vacuum Present : 0FF Estimated Water Fill Rate: 0.0 (liters per minute) Vacuum Turn On Time: (seconds) Monitoring of water/waste is disabled while using this diagnostic. F1: WATER ON/OFF F2: WASTE ON/OFF F3: F4:VACUUM ON/OFF F5: F6: F7: F8: Vacuum control signal is now ON. 06/25/23 18:15 Supplies Waste Full:ON Vacuum control signal is now ON, Automatic monitoring of water/waste is re-enabled. F4: SYSTEM CHECK SYSTEM PREPARATION MENU F1: INVENTORY F2: REAGENT PREP F5: TEMP MONITOR F6: SYS COUNTERS F7: PUMP PRIME PUMP PRIME CYCLES: 3 IMT SAMPLER REAGENT 1 REAGENT 2 DIL PUMP REAGENT 3 F1: PRIME WATER F2: SAMPLER F3: REAGENT 1 F4: REAGENT 2 F5: IMT DIL PUMP F6: HM ... F7: REAGENT 3 Pump prime in progress. Press (Enter) to cancel. MAINTENANCE LOGS Disclaimer Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Siemens Healthcare GmbH 2023
- Dimension
- dimension
- chemistry
- probe
- drain
- probe drain
- maintenance
- clean drains
- reagent
- reagent probe drains
- clean reagent probe drains