Cleaning the IMT System on the Dimension® Family of Integrated Chemistry Systems
Learn how to clean the IMT system on the Dimension® Family of Integrated Chemistry Systems.
Monthly cleaning of the IMT system Tubing and IMT port with bleach destroys sensitive electrode membranes in the Quicklight Integrated Multisensor. With the instrument in standby, access the IMT System Clean screen from the operating menu by selecting F4 System Prep, F3 IMT, F8 IMT Sysclean. From the IMT System clean screen, select F3 Open Port. Dispense approximately 1 milliliter of undiluted bleach into the IMT port 10 times. Allow the port to empty completely before each subsequent addition of bleach. Next, dispense approximately 1 milliliter of water into the IMT port 10 times. Allow the port to empty completely before each subsequent addition of water. Select F3, close port, Fill the port with undiluted bleach. And select F8 Start Soak. The fluid sits inside the tubing for two minutes. When the soak is complete. The system automatically primes standard A. When this process is completed. Fill the port with fresh serum or plasma free of hemolysis, ictorus or lipemia. Select F8. Start soak. Make sure the fluid is flowing in the tubing. The fluid sits inside the tubing for two minutes. When the soak is completed. The system Prime standard a select F7 change sensor. Next, follow the operator's guide instructions to replace the Quicklight Integrated Multisensor. The sensor must be replaced after bleaching the IMT system. Select F7 Change Sensor. The screen updates with the maximum assays and time for the new Quicklight sensor. Enter the new lot number for the new sensor. Select F8 Store Change. After the sensor is replaced, the system will automatically prime the salt bridge solution and display the IMT condition and dilution check screen. Perform the dilution check procedure. For additional information on this procedure, refer to the Operators guide Replacing IMT Consumables, running a Dilution Check. If documenting using the paper logs from the Dimension System Operators Guide, document the maintenance in the weekly slash, monthly maintenance log sheet or instrument log sheet, or if documenting using the electronic logs in the dimension system software, document the maintenance in the instrument log screen.
652 2976 648 1393 1000 647 Cleaning the IMT System on the Dimension® Family of Integrated Chemistry Systems SIEMENS Standby Sampler Idle IMT OK 09/14/22 21:23 Dimension Exit labmanager Enter en OPERATING MENU STAT Status PLEASE SELECT AN OPTION USING THE FUNCTION KEYS: Sample Alert At any time, press HELP to get help, or press EXIT to exit a program. Supplies QC Alert Calib Alert To Shut down or Change Operator, press Sign Out or Alt+EXIT. Run Home F1: ENTER DATA F2: SAMPLE STATUS F3: TEST RESULTS F4: SYSTEM CHECK F5: PROCESS CTRL F6: SYSTEM CONFIG F7: DIAGNOSTICS F8: SEGMENT STATUS 01/12/23 07:43 SYSTEM PREPARATION MENU F1: INVENTORY F2: REAGENT PREP F3: IMT F5: TEMP MONITOR F6: SYS COUNTERS F7: PUMP PRIME 06/12/23 14:20 IMT SETUP MENU F1: CHANGE CONSUM F2: CALIBRATION F3: ALIGN/PRIME F4: COND/DILCHK F5: IMT OVERVIEW F6: REVIEW DILCHK F7: IMT CONFIG F8: IMT SYS CLEAN IMT Paused IMT SYSTEM CLEAN IMPORTANT! AFTER PERFORMING THE BLEACH AND CONDITION PORTION OF THIS PROCEDURE, YOU MUST REPLACE THE QUIKLYTE SENSOR. Bleach will ruin or seriously degrade a new sensor. 1. Bleach Port: Press F3: OPEN PORT. Dispense 1 ml of undiluted bleach 10 times into IMT port, followed by 1 ml of water 10 times to rinse. Press F3: CLOSE PORT. 2. Bleach Tubing: Fill port with undiluted bleach. Press F8: START SOAK 3. Condition Tubing: Fill port with serum or plasma. Press F8: START SOAK 4. Change Sensor: Press F7. F1: F2: F3: OPEN PORT F4: F5: F6: F7: CHANGE SENSOR F8: START SOAK Rotary Valve IMT Port Rotary F2 Rotary V 06/12/23 14:30 F3: CLOSE PORT IM Interfac IMT System Soaking for two minutes ... 06/12/23 14:24 06/12/23 14:42 IEMENS QuikLYTE ElectroLYTE EMENS 2CD891 05143 06/12/23 16:41 CHANGE IMT CONSUMABLES Std A Std B Flush Salt Soln. Diluent Sensor Assays Remaining: 1957 LOT: QuikLYTE(tm) Integrated Multisensor Time Left: 5 days Standard A Time Left: 21 days Diluent Time Left: 21 days IMT bleach/conditioning soak due in 30 days Replace consumable, press the appropriate function key to update the count, and then store changes. F1: CHANGE STD A F2: CHANGE STD B F3: CHANGE FLUSH F5: CHANGE SALT F6: CHANGE DILUENT F8: STORE CHANGES Enter Sensor lot ID or ENTER to retain the current lot - 2CD891 5 days 2CD891 Priming Salt Bridge Solution. Processing IMT Not Calib IMT CONDITION AND DILUTION CHECK Select desired function. Place solutions in the designated cup positions: (When starting a new sensor cartridge, select F4: COND & DILCHK) Start at Position: A1 Solution/Sample Cup Number Conditioning Sample (Serum or Plasma sample) Dilution Check Sol. F1: CONDITION F2: DILCHK F3: F4: COND & DILCHK F7: F8: 06/12/23 17:48 Start at Position: A1 New cup(s) have been assigned. F5: START F7 MAINTENANCE LOGS Disclaimer Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Healthineers
- dimension
- Dimension
- IMT
- imt
- maintenance
- clean imt
- clean imt system