Cleaning the Entry and Exit Queues on the Atellica® Direct Load Job Aid Version 1.26 and Above
This job aid provides instructions on how to perform the as needed maintenance task of cleaning the Atellica® Direct Load entry and exit queues.
Cleaning the Entry and Exit Queues on the Atellica® SIEMENS Healthineers Direct Load Job Aid Cleaning the DL Entry and Exit Queues 1. Stop the DL: A. In the System Status area, select the appropriate DL Module B. Select Stop if the system is still processing samples, select the , option to wait for processing to stop or stop immediately 2. Remove debris from the DL entry and exit queues 3. Wipe the lanes with lint-free tissue dampened with a 10% bleach solution 4. Wipe the lanes with lint-free tissue dampened with reagent water 5. Dry the lanes thoroughly with clean lint-free tissue 6. Start the DL by selecting the Resume icon in the System Status area SIEMENS .: Healthineers Atellica Solution 6 LabManager DL Processing System Connected to PCC CH-01 Current Status Tool AF030672 AF030675 6 Unlock Front Cover Unlock Back Cover Service Diagnostics Template Effective Date: 10 Aug 2021| HILS 2223 1 Effective Date: 1-Nov-2021| HOOD05162003232996 Unrestricted © Siemens Healthineers, 2021 SIEMENS Healthineers Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. © Siemens Healthcare GmbH 2021 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens-healthineers.com 2 Effective Date: 1-Nov-2021| HOOD05162003232996 Unrestricted © Siemens Healthineers, 2021
- attelica
- attellica
- atelica