This video provides instructions on how to perform the monthly maintenance task of cleaning the Atellica® Direct Load fan filter. The fan filter is a customer replaceable part.
The Atellica direct load fan filter must be cleaned monthly to ensure proper airflow through the direct load and prevent overheating. The direct load fan filter is located in the lower area of the back direct load panel. To clean the fan filter, first grasp the top of the fan filter and pull it out of the slot. Inspect the filter. If the filter is damaged or unable to be cleaned, discard per laboratory protocol and replace with a new filter that was provided in the customer parts kit. If the filter is able to be cleaned, rinse the filter with tap water until clean, then tap the filter to remove excess water and let air dry. Complete the procedure by reinstalling the fan filter. The spare filter from the customer parts kit can be used while the cleaned filter dries. Slide the filter into the panel, making sure the plastic grid is facing away from the direct load.