Cios Spin Measurements and Annotations Job Aid
This job aid provides an overview of the measurement and annotation tools that enable measuring distances, lengths and angles of relevant structures or adding annotations to images.
SIEMENS Healthineers Measurements and Annotations Job Aid 3D ealthineers Clos Spin Cios Spin CION abge Template Effective Date: 10 Aug 2021 | HILS 2223 Job Aid Effective Date 12 APR 2022 | HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Table of Contents AAA_Cios_Alpha 10 0067 Introduction 3 01-01-1932 Vasc AAA Displaying the Evaluation Functions 4 Control Panel: General Steps in the Measure Annotate Menu 5 XYZ 1 Distance = 18.5 mm Calibration Object 10 Control Panel: Perform Calibration 11 Control Panel: Distance Drawing Mode 12 Orbital 0º Angular 0º ‹ ‹ Control Panel: Angle Drawing Mode 14 < 0.1 mGy < 0.01 uGy*m2 Control Panel: Annotate Images 15 00:00:00 hh:mm:ss Control Panel: Setting the Side Identification 18 B Evaluation in Enhanced Mouse Mode 20 Single Image Fluoro Enhanced Mouse Mode: General Steps 21 6 x 976 Enhanced Mouse Mode: Perform Calibration 23 SUB lask#1 NAT 2 SUB 2 Enhanced Mouse Mode: Measure the Distance and Angle 24 Service switch S3. ON ige 1.00 v: 4095 ndard Enhanced Mouse Mode: Annotate Images 25 Save Images During Postprocessing 27 2 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Introduction With the measurement and annotation tools, you can AAA_Cios_Alpha 30-30 VA 10 0067 2-2018 01-01-1932 3 AM measure distances, lengths and angles of relevant structures Vasc AAA or add annotations to your images. During image evaluation, the left monitor switches Distance = 18.5 mm automatically to 1x1 layout. When the evaluation is finished the original monitor layout is restored. Graphical evaluations can only be performed on a single image at a time. The graphic objects are saved in the image Orbital 0º automatically. Angular 0º V < 0.1 mGy < 0.01 uGy*m2 00:00:00 hh:mm:ss Please note: Measurements are not possible for images imported from other modalities and images from external video sources (CAP images). Single Image Fluoro 76 x 976 SUB Mask#1 NAT 2 SUB 2 Service switch S3. ON dge 1.00 4095 ndard C: 1600 3 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Displaying the Evaluation Functions The image evaluation tools can be accessed in two different ways. On the following slides, we will first look at how you use the evaluation functions via the control panel. From slide 20, you will learn more about the enhanced mouse mode. Measure Annotate Menu on Control Panel Enhanced Mouse Mode on Left Monitor You can use one of the system's control panels (touch user interface). If you are working on the monitor trolley, you can also switch to the There you will find the evaluation tools within the Measure Annotate enhanced mouse mode and perform all evaluation functions directly menu. on the left screen with the mouse. You can only use the control panel to enter text. -x- Orb.0º Set 1st Delete X Store point point AAA_Cios_Alpha 12-2018 Ang.0º XYZ position 01-01-1932 XYZ . X VaSC AAA 62 68 Calibrate Delete Phrase 1 selected uGym2 KV XYZ 00:00:31 17 Distance = 18.5 mm Free text XYZ hh:mm:ss mA X * XYZ Collimator Laser Reset timer Line Angle Text Delete all Orbital 0º Angular 0º Fluoro Single 3D R RJA XYZ < 0.1 mGy image Scan Rotate Live graph Zoom Pan Measure Reset all < 0.01 uGy*m2 Flip overlay Shutter Annotate 00:00:00 hh:mm:ss Sub Road DCM REF Brightness Contrast Overview Hold Cine reference replay Previous Next Single Medium Auto Image Fluoro Dose Metal Tech lock A + B correction Image Target enhance Pointer Transfer Store Store cine Print Export Mag 0 15 p/s Magnify Motion Noise filter Pulse rate PREPARATION PROCESSING EXTERNAL K Close pat. Keyboard Service switch S3: ON 4 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Control Panel: General Steps in the Measure Annotate Menu Open the Measure Annotate Menu 3 1. Open the Processing task card. Orb. 0º × Set 1st Delete point point 2. Press the Measure Annotate button. Ang.0º Store position XYZ . X 3. The Measure Annotate menu is opened. 62 68 Calibrate Delete Phrase 1 selected Gym2 KV 00:00:31 17 Free text R hh:mm:ss mA * XYZ Collimator Laser Reset timer Line Angle Text Delete all Fluoro Single 3D R RIA image Scan i KXYZ Rotate Live graph Zoo 2 Measure Flip overlay Shu Annotate Reset all Sub Road DCM REF + Brightness Hold Cine Contrast Overview reference replay Previous Next Medium Auto Dose Metal correction Tech lock A + B Image Target enhance Pointer Transfer Store Store cine Print Export Mag 0 15 p/s Magnify Motion Noise filter Pulse rate PREPARATION PROCESSING EXTERNAL Close pat. Keyboard 1 5 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Control Panel: General Steps in the Measure Annotate Menu Move the Mouse Pointer - Orb. 0º × Set 1st Delete Use the mouse pointer to set the position in the image where Store point point you want the processing of graphic objects to take place (such Ang.0º position 62 XYZ . X Calibrate Delete Phrase 1 as insert, select). 68 selected uGym2 kV Use the arrow buttons to move the mouse pointer in the 00:00:31 17 image. hh:mm:ss mA IIIII Free text L R Vertical arrow buttons: Move up/down * XYZ • Collimator Laser Reset timer Line Angle Text Delete all • Horizontal arrow buttons: Move left/right Diagonal arrow buttons: Combined horizontal and vertical Fluoro Single 3D • image Scan -. Rotate Live graph Zoom Pan Measure movement Flip overlay Shutter Annotate Reset all Sub Road DCM REF + Brightness Contrast Overview Hold Cine reference replay Previous Next Or – Medium Auto – Metal A + B Move the mouse pointer with the mouse on the monitor Dose correction Tech lock Image Target enhance Pointer Transfer Store Store cine Print Export trolley. Mag 0 15 p/s Magnify Motion Noise filter Pulse rate PREPARATION PROCESSING EXTERNAL Close pat. Keyboard 6 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Control Panel: General Steps in the Measure Annotate Menu Select a Graphic Object - Orb. 0º × Set 1st Delete To move, edit, or delete an existing graphic object, you must Ang.0º Store point point position first select the object. XYZ . X 62 68 Calibrate Delete Phrase 1 To do that, move the mouse pointer with the arrow buttons or selected uGym2 kV the mouse on the monitor trolley to the graphic object. 00:00:31 17 When an object is selected, the reference points (vertices, hh:mm:ss mA IIIII Free text L R center point) are highlighted. * XYZ Collimator Laser Reset timer Line Angle Text Delete all Fluoro Single 3D image Scan -. Rotate Live graph Zoom Pan Measure Flip overlay Shutter Annotate Reset all Sub Road DCM REF + Brightness Overview Hold Cine Contrast reference replay Previous Next Medium Auto Dose Metal correction Tech lock A + B Image Target enhance Pointer Transfer Store Store cine Print Export Mag 0 15 p/s Magnify Motion Noise filter Pulse rate PREPARATION PROCESSING EXTERNAL Close pat. Keyboard 7 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Control Panel: General Steps in the Measure Annotate Menu 5 Move a Graphic Object End line 1. Select the graphic object. 2 2. Press Set 1st point. The graphic object is linked to the -X- × Set 1st Delete mouse pointer. Orb. 0º Store point point Ang.0º position 3. Position the mouse pointer (+) to the desired starting 62 XYZ . X 68 Calibrate Delete Phrase 1 position. 6 elected Gym2 KV 4. Move the graphic object to the required position with the 00:00:31 17 arrow buttons or mouse. hh:mm:ss mA Free text L R 5. Press End line. The link between the graphic object and * 1 + 4 XYZ Collimator Laser Reset the mouse pointer is removed. timer Line Angle Text Delete all 6. The Calibrate button can now be selected. The button is Fluoro Single 3D XYZ image Scan -. dimmed until you draw a line and is only highlighted if Rotate Live graph Zoom Pan Measure Flip overlay Shutter Annotate Reset all the + is over the line. Sub Road DCM REF + Brightness Overview Hold Cine Contrast reference replay Previous Nex Medium Auto Dose Metal correction Tech lock A + B Image Target enhance Pointer Transfer Store Store cine Print Export Mag 0 15 p/s Magnify Motion Noise filter Pulse rate PREPARATION PROCESSING EXTERNAL Close pat. Keyboard 8 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Control Panel: General Steps in the Measure Annotate Menu Delete Selected Graphic Objects Delete All Graphic Objects 1. Select the graphic object. Press Delete all. 2. Press Delete selected. All graphic objects in the image are deleted. The selected graphic object is deleted. x- × -x- × O Set 1st Delete Set 1st Delete point point point point XYZ . X XYZ . X Calibrate Delete Phrase 1 Delete Phrase 1 selected Calibrate selected Free text L R Free text L R XYZ XYZ Line Angle Text Delete all Line Angle Text Delete all 9 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Calibration Object If you perform evaluations with distance measurements, you must AAA_Cios_Alpha AAA Cios Alpha Spin 30130 VAJO 0067 12-2018 calibrate the image. 01-01-1912 2:51 AM 01-01-1932 Vasc AAA For a calibration to be performed, there must be a calibration object in the image. For increased accuracy, it is recommended to use a calibration object that is as large as possible, but still can be displayed completely. The edges of the calibration object must be clearly recognizable, and its dimensions must be known. Orbital 0º Please note: The calibrating accuracy decreases if the calibration Angular 0º object is not located in the same plane as the measured object. < 0.1 mGy < 0.01 uGy*m2 00:00:00 hh:mm:ss Single Image Fluoro 8 61.4 HV 76 1 976 0,60 mAy SUB D: 0.015 k#1 Mag 0 Service switch 53: ON 1.00 1 16/174 3242 Vasc Standard C: 1340 10 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Control Panel: Perform Calibration Perform Calibration X 1. First, draw a distance line along the calibration object. Set 1st Delete point point Press the Line button to activate distance drawing mode. Be careful to mark the starting and end points 2 XYZ . X Calibrate Delete Phrase 1 with as close to pixel accuracy as possible in order to selected achieve optimal measuring accuracy. For more information regarding distance drawing Free text L R mode, please see the next slide. 2. When your line is complete, press Calibrate. XYZ The control panel switches to keyboard mode. 1 Line Angle Text Delete all 3. Enter the distance (mm) and confirm with Enter. A length scale is displayed in the image and is applied to the remaining images of the series. Distances already drawn in are updated. 11 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Control Panel: Distance Drawing Mode Using a distance line, you can measure the distance between two points in an image. 3 End line 1. Press the Line button to activate the Distance drawing mode. 2. Use the arrow buttons to move the mouse pointer to the starting point of the line, and press Set 1st point. 2 × Set 1st Delete point point Or click the starting point with the left mouse button. The button changes to Set last point. XYZ . X Calibrate Delete Phrase 1 3. Use the arrow buttons to move the mouse pointer to the last selected point of the line and press End line. Or click the last point with the left mouse button. Free text L R The length (mm) and the sequential number of the graphic XYZ object are displayed at the end point of the line. 1 Line Angle Text Delete all 12 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Control Panel: Distance Drawing Mode Delete Points × If the starting point or end point has been placed incorrectly, it can be Set 1st Delete point point deleted immediately after being drawn. XYZ . X 1. To delete a point, select Delete point. The last point set is deleted Calibrate Delete Phrase 1 from the image. selected 2. If the starting point is deleted, add a new starting point by pressing Set 1st point to draw the line again. Free text L R XYZ Line Angle Text Delete all 13 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Control Panel: Angle Drawing Mode You can define an angle of two lines by drawing the legs of the angle on the image. 3 End line 1. Press the Angle button to activate the Angle drawing mode. 2. Draw the first line in the image by setting the starting point and - end point. 2 × O Set 1st point Delete point 3. Draw the next line in the image by setting the starting point and end point. XYZ . X Calibrate Delete Phrase 1 selected Now, the angle is set. The drawn angle (in °) and the sequential number of the graphic object are displayed at the end points of the R legs. Either an inner or outer angle is indicated depending on the Free text L R starting point and end point of the second leg. 1 XYZ Line Angle Text Delete all 14 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Control Panel: Annotate Images Enter Free Text --- X You can annotate interesting or anomalous areas in the image by entering 2 Set 1st point Delete point free text. XYZ . X Delete Phrase 1 1. Press the Text button to activate the Text drawing mode. Calibrate selected 2. Move the mouse pointer to the required position for the annotation 3 with the arrow buttons or mouse. Free text L R 3. Press the Free text button. The control panel switches to keyboard XYZ mode. 1 Line Angle Text Delete all 4. Enter the text and confirm with Enter on the keyboard. The text is X inserted at the position of the mouse pointer. ESC F1 F2 F3 F4 F5 F6 F7 F8 F9 F10 F11 F12 1 2 4 5 6 7 8 9 O = Back space Tab W e r y u i p O - 4 Caps Lock a s d f g h j k Shift C V b n m Ente N Ctrl Alt 15 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Control Panel: Annotate Images Insert Predefined Text - × O Set 1st Frequently used texts can simply be selected from a template once they point Delete point have been created as such. The predefined texts are created on the monitor XYZ . X trolley (please refer to page 26 Creating and deleting predefined texts for Calibrate Delete Phrase 1 selected more information). Select the required text from the selection list. The text is inserted at the Phrase 1 R position of the mouse pointer. My note Line Angle Delete all Pelvis Rotate Flip Sub / Native PeakOP Live graph overlay Reset all Test phrase 1+ REF LO + Brightness Select Hold Contrast mask Overview reference Next A + B Image Pixel shift Target Landmark Pointer Transfer enhance Store -TITIL Export PREPARATION PROCESSING EXTERNAL K Close pat. Keyboard 16 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Control Panel: Annotate Images Change Text × 1. Select the graphic object with the text. Set 1st point Delete point 2. Press the Free text button to change the text using the XYZ . X keyboard. Delete Phrase 1 Calibrate selected Or – – Select a new text from the selection list. Free text L R XYZ Line Angle Text Delete all 17 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Control Panel: Setting the Side Identification Add Side Identification 2 - Set 1st Delete point X By identifying the side of the patient, you avoid a mistake in image point orientation. The side identification is applied automatically to the remaining images of the series. XYZ . X Calibrate Delete Phrase 1 selected 1. Switch on the Text drawing mode, if required. R 2. Move the mouse pointer to the required position for the identification (such as the edge of the image) using the arrow Free text L R 3 buttons on the control panel or the mouse on the monitor XYZ trolley. Line Angle Text Delete all 3. Press the button for the correct patient side. The selected identification L or R is inserted at the position of the mouse 1 pointer. 18 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Control Panel: Setting the Side Identification Change Side Identification - Set 1st X Press the L or R button for the correct patient side. point Delete point The original side identification is deleted. XYZ . X The selected identification L or R is inserted at the position of Calibrate Delete Phrase 1 selected the mouse pointer. R Free text L R XYZ Line Angle Text Delete all 19 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Evaluation in Enhanced Mouse Mode Activating Enhanced Mouse Mode AAA_Cios_Alpha AAA Cios Alpha tips Spin 30:30 VAJD OUGT 12-2018 01-01-1932 01.01.1912 00 2:51 AM 1. XYZ Select this button on the left monitor. Vase AAA L 2 The evaluation menu with the mouse is displayed in the upper X left of the left monitor. 2. Select the button for the required function. Enhanced mouse mode is activated and the control panel switches to keyboard mode. Orbital 0º Angular 0º < 0.1 mGy :0.01 pGy*m2 Turning off Enhanced Mouse Mode 00:00:00 hh:mm:ss Click the button of the activated function. Or end keyboard mode on the control panel by selecting Single × Image Fluoro the X. 61.4 AV 176 x 976 1 0.60 mAs SUB D: 0.015 Mask#1 Service switch 53. ON 1,00 ₩ 3242 Standard C: 1340 20 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Enhanced Mouse Mode: General Steps Selecting an Object, Moving an Object and AAA_Cios_Alpha AAA Cios Alpha tips Spin 30:30 VAJD OOGT 12-2018 01-01-1932 01-01-1912 00 2:51 AM Changing Dimensions Vasc AAA You must select an existing graphic object to process it. L 1. Click the Selection button. +++ 2. Select the graphic object. To move the graphic object, press the left mouse button, and drag and drop the center point to the required position. Orbital 0º To change the dimensions of the graphic object, press the left Angular 0º mouse button and drag & drop the vertices in the required < 0.1 mGy direction. :0.01 pGy*m2 00:00:00 hh:mm:ss XYZ Single Image Fluoro ----- 1 XYZ 61.4 AV 176 x 976 0.60 mAs SUB D: 0.015 Mask#1 X 1,00 ........ Service switch 53: ON ₩ 3242 - Standard C: 1340 21 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Enhanced Mouse Mode: General Steps Healthineers Delete All Graphic Objects Delete Selected Graphic Objects To delete all graphic objects in the image, select the highlighted button 1. To delete individual graphic objects, select the graphic object. on the left monitor. 2. Press the Backspace button on the control panel to delete the selected object. X XYZ ESC F1 F2 F3 F4 F5 F6 F7 F8 F9 F10 F11 F12 1 2 3 4 5 6 7 8 9 0 Back = space O Tab W e b r t y u O XYZ p × Caps Lock a s d h k ........ g - . - Shift Z X c b n m Enter - < Ctrl Alt 22 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Enhanced Mouse Mode: Perform Calibration 1 2 1. Draw a distance line (see following page) along the calibration XYZ object and select it. . 2. Then click Calibrate. 3. The Calibrate dialog box is displayed. Enter the distance in XYZ ........ ........ × ......... ......... arbitrary units (e.g. in mm) and select OK. ..... A length scale is displayed in the image and is applied to the AAA_Cios_Alpha AAA Cios Alpha Cios Spin 30-30 VA10 006 12-2018 remaining images of the series. Distances already drawn in 01-01-1932 01-01-1912 00M 2:51 AM are updated. Vasc AAA L Please note: The current calibration cannot be deleted or stored. You can only overwrite the current calibration by setting up a new x calibration. Orbital 0º Calibrate x Angular 0º Length 24 3 mm < 0.1 mGy < 0.01 pGy*m2 Cancel 00:00:00 hh:mm:ss 23 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Enhanced Mouse Mode: Measure the Distance and Angle Measuring the Distance 1. Select the Line button. Line Button Angle Button 2. In the image, select the starting point of the line and the XYZ desired end point to complete the distance line. Measuring an Angle 1. Select the Angle button. 2. Draw the first line in the image by setting the starting point and end point. XYZ 3. Repeat the process for the second line to complete the ......... ........ angle. × ......... 24 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Enhanced Mouse Mode: Annotate Images To add annotations, do the following: XYZ 1. Select the Annotation button. The input dialog box for text and 1 comments is displayed. 2. To set the side identification, select the appropriate patient side in the dialog box and the appropriate location in the image. 3. To insert annotations, select the annotation from the Text selection XYZ ......... list. You can also select the text field of the Text selection list and × ......... ........ overwrite the displayed text. Then click the required position in the image. Text 4. To add a comment, select the text from the Comment selection list. 2 L R Or click the text field of the Comment selection list and overwrite the displayed text. Comments always appear centered at the bottom edge 3 xyz of the image and are transferred automatically to all images of the Add to predefs. Del from predefs. series. XYZ Comment 4 Zyx Add to predefs. Del from predefs. XYZ × 25 ......... Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Enhanced Mouse Mode: Annotate Images Creating and Deleting Predefined Texts XYZ You can save frequently used texts and comments as templates and select them from the list as needed. 1. To create a predefined text, enter the text in the text field of the corresponding selection list. XYZ 2. Click the Add to predefs. button. The text is added to the list as a ......... × ......... new entry. Up to 20 entries each can be created in the Text and ........ Comment selection lists. Text 3. To delete a predefined text from the selection list, select the entry you no longer require. Then, click the Del from predefs. button. L R 1 xyz 2 Add to predefs. Del from predefs. 3 Comment zyx 2 Add to predefs. Del from predefs. 3 26 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 Measurements and Annotations – SIEMENS Healthineers Save Images During Postprocessing After completing postprocessing of a prior examination, all -X O changes in the image are saved when you close the Orb. 0º X Set 1st Delete patient. To close a patient, select the Close pat. button. Ang.0º Store point point position To save the displayed image with its changes, press the 62 68 XYZ . X Phrase 1 Calibrate Delete Store button on the control panel. KV selected Gym2 If you want to save intermediate states, manually save the 00:00:31 17 hh:mm:ss mA Free text L R image in the required processing phase. The image is saved as a copy of the original in the corresponding study. * XYZ Collimator Laser Reset timer Line Angle Text Delete all Fluoro Single 3D R RJA EXYZ image Scan -. Rotate Live graph Zoom Pan Measure Flip overlay Shutter Annotate Reset all Sub Road DCM REF + Brightness Overview Hold Cine Contrast reference replay Previous Next Medium Auto Dose Metal A + B correction Tech lock Image Target enhance Pointer Transfer Store Store cine Print Export Mag 0 15 p/s Magnify Motion Noise filter Pulse rate PREPARATION PROCESSING EXTERNAL K Close pat Keyboard 27 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022 SIEMENS Healthineers Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. © Siemens Healthcare GmbH 2022 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens-healthineers.com 28 Effective Date: 12 APR 2022| HOOD05162003266153 Unrestricted © Siemens Healthineers, 2022
- ciosspin
- c-arm
- measurement
- annotation
- mobile c-arms
- carm