Changing Field of View size on the SOMATOM go. system
Job aid demonstrating the steps to change Field of View size on
the SOMATOM go.Sim and go.Open Pro system.
Changing Field of View size on the SOMATOM go. system SOMATOM go.Sim, go.Open Pro (VA40) Depending upon the scanner bore size, there are different default field of view (FoV) sizes. On the go.Sim and the go.Open Pro, the default FoV is 600mm with a maximum HD FoV of 850mm. Sometimes all the desired anatomy is not visualized in the default FoV. There are two workflow processes to adjust the FoV: 1) before the reconstruction or 2) adding a reconstruction if anatomy is not seen or to reduce the FoV in the previous reconstruction. Change the FoV before the reconstruction 1. You may determine after the topogram is acquired that the FoV is not appropriate. On the Reconstruction box, right-mouse click and select Show Parameters. change, Fov "5/30/1973, 50Y Recon Ranges 5/30/2023 2.26.38.22 PM 2 IMA 1 TOP 201 TP -1.262.0 Show Parameters Delete Repeat Copy All Recon Parameters Copy Recon Volume Parameters Paste All Recon Parameters Paste Recon Volume Parameters Create Image with Reference Lines Send to CT View&GO Edit Result in CT View&GO 2. On the Recon Parameters toolbar, if the FoV needs to be opened wider, check the HD FoV box On. Note: For smaller field-of-view sizes, there is no need to check the HD FoV box. Recon Favorites General Recon Recon&GO Image Impression Recon Box Physio Recon Auto Tasking Inline Options Scan/Recon Image FAST Planning Matrix Width Height Center Center Y Mirrorin Orientation Image Order FAST Planning Width Matrix [mm] [mm] [mm] HD FOV [mm] Pelvis 1.50 Qr40 A3 Axial Head to Feet " Pelvis - Iliac Crest to Troc ... Full Square 500.00 500.00 0.00 On 3. A pop-up box will appear with an artifact warning. Select OK. HD FOV × ? Reconstruction with HD FoV can lead to image artifacts. OK Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2023 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 07/10/2023 | HOOD05162003362601 Healthineers 4. On the Recon Parameters toolbar, look for the Width and/or Height box. Type in the new FoV. Recon Favorites General Recon Recon&GO | Image Impression Recon Box Physio Recon Auto Tasking Inline Options Scan/Recon Image Image Order FAST Planning FAST Planning Matrix Width Height Center X Center Y Mirrorin Orientation Width Size Matrix [mm] [mm] [mm] [mm] HD FOV Pelvis 1.50 Qr40 A3 Axial Head to Feet " Pelvis - Iliac Crest to Troc ... Full 512 700.00 = 700.00 0.00 0.00 None Note: In this example, the matrix is set to square so you can type in the Fov in either box and they will both update to the desired size. If you are using an Inline Result, in order to type in a number in the Width and Height boxes, you must first turn it to None. This can be found in either Inline Results or under the Inline Options tab. 5. The new FoV will then be displayed on the topogram. Complete the scanning process. Change the FoV after the reconstruction is complete 1. Right-mouse click on the previous reconstruction and select Repeat. change, FoV SOMATOM go Open Pro fow syngo CT VADA Recon Ranges 5/30/1973, 50Y 5/30/2023 2:49:47.96 PM 3 IMA 118 SPI 301 TP -1,437.0 Show Parameters Delete Repeat Copy All Recon Parameters Copy Recon Volume Parameters Paste All Recon Parameters Paste Recon Volume Parameters Create Image with Reference Lines Send to CT View&GO Edit Result in CT View&GO Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2023 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 SIEMENS Job Aid Effective Date: 07/10/2023 | HOOD05162003362601 Healthineers Page 2 of 4 2. Right-mouse click on the repeated recon box and select Show Parameters. TOM go.Open Pro syngo CT VA40A Recon Ranges HFS Topograms Pelvis Show Parameters Delete Repeat Copy All Recon Parameters Copy Recon Volume Parameters Paste All Recon Parameters Paste Recon Volume Parameters ing Create Image with Reference Lines Send to CT View&GO Edit Result in CT View&GO Cancel 3. As with changing the FoV before the reconstruction example above, follow steps 3-5 to turn on the HD FoV (for larger field sizes). Note: For smaller field sizes, there is no need to check the HD FoV box. 4. Select Recon All to reconstruct the new parameters. Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2023 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 SIEMENS Job Aid Effective Date: 07/10/2023 | HOOD05162003362601 Healthineers Page 3 of 4 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2023 Siemens Healthineers Headquarters Local Contact Information Legal Manufacturer Siemens Healthcare GmbH Siemens Medical Solutions USA, Inc. Siemens Shanghai Healthcare Sector Henkestr. 127 40 Liberty Blvd. Medical Equipment Ltd. 91052 Erlangen Malvern, PA 19355-9998 USA 278 Zhou Shu Road Germany Telephone: +1-888-826-9702 Shanghai 201318 Phone: +49 9131 84-0 Customer Care Telephone: P.R. China https://www.siemens-healthineers.com +1-800-888-7436 Telephone: +1 847 304 7700 Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2023 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 SIEMENS Job Aid Effective Date: 07/10/2023 | HOOD05162003362601 Healthineers Page 4 of 4
- fov
- missing anatomy