Atellica® DT 250 Order and Load a Calibration Video

This video demonstrates how to order and load a calibration onto the Atellica DT 250 analyzer.

In this video, we will demonstrate how to order and load a calibration onto the Atelica DT-250 Analyzer. Let's begin by learning how to order a calibration. Start by selecting the calibration icon in the menu bar. Select the calibration button and then select New. From the new calibration window. Select a calibrator set from the Calibrator set drop down. Then select the test or tests you want a calibration to perform. To select all tests, select the blue box in the column header. Additionally, if you would like to run a test multiple times for precision testing, utilize the Replicate column. When finished, select OK to confirm your input. Now let's navigate to the sample tray screen to load our calibration. Select the sample tray icon in the menu bar. Select an available sector from the sector list, and then select Place Sample from the Place Sample window. Use the arrows to locate and select the Calibrators tab. Then select the calibration we ordered. Select Place All to automatically assign the calibration level sequentially to the available sector positions. Or you can manually assign the calibrator levels to sector positions by selecting positions from the Positions field and then selecting Place. In this example, we will select Place All. Select Exit to return to the sample tray screen. Our place calibrators are indicated by the green positions 1-2 and three in the displayed sector. Next, select Secondary to process the calibration from sample cups. The ring around the sector positions indicates Secondary mode. Pipe out the calibrator levels into the sample cups and place them in the correct sector positions. When complete, select Place Sector. The sample tray drives the first available sector position to the sample tray loading area. Slide open the sample tray loading area and insert the sector. Once the sector is properly secured and the sliding cover is closed, select OK Finally, navigate to the main screen by selecting the main screen icon in the menu bar. From here, select the Start button to begin analysis.

01:35 567890 00:00 21:05 20260418 200 19 17 15 11 13 18 16 10 12 14 Master Template HILS2218 | Effective Date: 2 Jan 2025 Order and Load a Calibration Atellica® DT 250 SIEMENS Atellica DT 250 Calibration Reaction Pending Autoanalyzer Activity Tests in Progress Calculated Pending Estimated Time to Completion 9:13 PM Method ID Sample ID Calibration pending acc. Run Options Next Deferred Dispensing Do Not Stop Operation Next Reading Amik Reagent blank pending acc. Remaining Tests Remaining Readings QC Pending System Messages Event Time Row Item Automatic Clot Calibration Maintenance In Tray Volume Calibrated oc # Tests # Available Cuvette Blank Lido System Flush Tip Cleaning Reaction Tray Full Fluid Levels Severity Leve Observation Aux. Off System Alerts Calibration Required +C15 Reagent Blank Required System Log 50 % 11 % 100 % +C30 +EX3 Connection Waste (S3) Sys. Water (S4) Cleaner +EX5 Offline Operating LIMS System Water Cleaner Auxiliary Pumps SIEMENS . Healthineers Reaction pending acc. Emit® Calibrator/Control Calibration Calculated pending acc. Pending Test Number Replicate Vial Lot# Expiration Concentration Units BENZ 200Q Y1 200 ng/mL Pending Calibration Acceptance Pending QC In Use Historic Maintenance Required Sector Position Cuvette blank Required System flush Required Set Tip cleaning Required New -F2 Delete - F3 Add Test - F5 Delete All New Calibration Calibrator Set Expiration Concentration AMPH 500Q BARB 200/3005 BARB 200Q BARB 300Q BENZ 200/300S BENZ 300Q COC 150Q COC 300Q Ok Cancel 2 v Atellica DT 250 Sample Tray Sector Sector operation Waiting Positioning Confirming Place Sector Remove Sector Empty Sector Place Sample Remove Sample Primary Secondary Calibrators Positions Emit® Calibrator/Control -4 Place All Exit Place Open Atellica DT 250 Main Calibration Pending Reagent Blank Pending OC Disclaimer Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc ., or its affiliates. All other trademarks are the property of their respective owners Unrestricted | Published by Siemens Healthineers AG | @ Siemens Healthineers AG, 2025 Siemens Healthineers HQ | Siemens Healthineers AG Siemensstr. 3 91301 Forchheim Germany Phone: +49 9191 18-0 siemens-healthineers.com

  • atellica
  • dt
  • 250
  • analyzer
  • system
  • overview
  • hardware
  • software
  • drug
  • testing