Atellica® DT 250 Monitor and Review Patient Results Video

This video demonstrates how to monitor and review patient results on the Atellica DT 250 analyzer.

In this video, we will demonstrate how to monitor and review patient samples on the Atelica DT-250 Analyzer. Begin by selecting the test icon in the menu bar. From the test screen, select Pending to view non processed or in progress sample tests. Now select Cumulative Historic. This tab allows you to search permanently stored results. Here you can import the file of the desired month and year of data. Now let's discuss how to accept a test result. Select Pending acceptance results under the Pending acceptance category weight for user confirmation. If the test of the method are set to manual acceptance and or the reaction is flagged, select a desired sample result and review any associated flags. You can confirm the result by either selecting Accept to accept the result, selecting Reject to reject the result, or selecting Rerun 1 to rerun the selected sample. The Done tab can be selected to review accepted and rejected results.

01:35 567890 00:00 21:05 7/1/2025 15 +5 +2 +3 +4 90 188 189 208 0.6 50.5 0.3723 131.0 0.55 0.6195 0.5 0.45 0.2455 0.4 0.0000 1.00 Master Template HILS2218 | Effective Date: 2 Jan 2025 Monitor and Review Patient Results Atellica® DT 250 SIEMENS . Healthineers Atellica DT 250 Tests Reaction Pending Autoanalyzer Activity Tests in Progress Calculated Pending Estimated Time to Completion 9:13 PM Method ID Sample ID Calibration pending acc. Run Options Next Deferred Dispensing Lid Do Not Stop Operation Next Reading Amik Reagent blank pending acc. Remaining Tests Remaining Readings QC Pending System Messages Event Time Row Item Automatic Clot Calibration Maintenance In Tray Volume Calibrated # Tests # Available Cuvette Blank Lido System Flush Tip Cleaning Reaction Tray Full Fluid Levels Severity Observation Aux. Off +BUP System Alerts Calibration Required +C15 Reagent Blank Required System Log 11 % 100 % +C30 +EX3 Connection Waste (S3) Sys. Water (S4) Cleaner +EX5 Offline Operating LIMS System Water Cleaner Auxiliary Pumps Reaction pending acc. Pending Sample Date Test Replicate Status Lot Number 12/09 13:36 METH150Q Test METH 150Q 1 In Progress Calculated pending acc. Acceptance Pending Calibration External Pending QC Calculated Done Maintenance Required Cuvette blank Required Cumulative Historic System flush Required Tip cleaning Required Sample Creation Date Last Name Patient csa b1 Samples Sample Type Reference class Collection Internal # csa b11 Other ADULT csa b12 csa b13 Status Result Units Flags csa b2 CSA - 7/1/2025 10:05:11 AM csa b21 csa b22 Accepted 57.6 ng/mL csa b23 65.7 ng/mL csa sp4001 61.1 ng/mL 69.4 ng/mL 70.7 ng/mL gENT NEG 541 MTX TDM 1 51 MTX TDM 2 52 MTX TDM 3 53 tdm 1 48 tdm 2 49 tdm 3 50 Print Export Cuvettes Sample Id Patient Id Low Limit Result High Limit Units Condition QC out of range METH150Q T METH 150Q Replicate: 1 - ng/mL First Negative ng/mL Positive Readings (sec. abs.) Comments Lot Number: u1 Measurement Absorbance Blank meas. Dil. factor 50 60 70 80 90 100 110 120 130 Time (sec) Accept - F2 Reject - F3 Rerun One - F4 Rerun Pos - F5 Rerun Diluted - F6 Refresh Table - F7 50 60 70 80 90 100 110 120 130 Disclaimer Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc ., or its affiliates. All other trademarks are the property of their respective owners Unrestricted | Published by Siemens Healthineers AG | @ Siemens Healthineers AG, 2025 Siemens Healthineers HQ | Siemens Healthineers AG Siemensstr. 3 91301 Forchheim Germany Phone: +49 9191 18-0 siemens-healthineers.com

  • atellica
  • dt
  • 250
  • analyzer
  • system
  • overview
  • hardware
  • software
  • drug
  • testing