Atellica® DT 250 Define a New Control Set Video
This video demonstrates how to define a new control set on the Atellica DT 250 analyzer.
In this video, we will demonstrate how to define a new control set on the Italica DT-250 Analyzer. Note that this procedure requires you to be logged in with supervisor privileges. New control sets must be defined when a new lot is available and new admissible ranges are defined. Begin by selecting the quality control icon in the menu bar. Let's start with defining A qualitative control set. Select the qualitative radio button. Select New. Define the control set ID. Enter the lot number and expiration date. These can be entered manually or with the optional barcode scanner. Double click the name of the test in the Quick Load field to add it to the control set. Enter the minimum separation absorbance value and the desired amount of replicates. Select OK to save. You have the option to import minimum separation absorbance values from an existing database. From Settings, select Data, Select Import and select QC Sets. Navigate to the location where they're stored, Select the ID you want to import, and select OK. Now let's see how to define a quantitative control set. Select the Quantitative radio button. Select New, define the control set ID, enter the lot number and expiration date. These can be entered manually or with the optional barcode scanner. Select OK to save.
567890 01:35 00:00 21:05 11% 50% 1/31/2026 0.040 0.081 0.032 8/28/2026 0.005 0.025 1/1/2027 4/17/2025 9/9/9999 9/30/2025 4/21/2028 300.0 270.0 330.0 7.5 5.0 10.0 450 300 600 Master Template HILS2218 | Effective Date: 2 Jan 2025 Define a New Control Set Atellica® DT 250 SIEMENS . Atellica DT 250 Main Reaction Pending Autoanalyzer Activity Tests in Progress Calculated pending acc. Estimated Time to Completion 9:13 PM Method ID Sample ID Next Deferred Dispensing Calibration pending acc. Run Options Lid Do Not Stop Operation Next Reading Amik Reagent blank pending acc. Remaining Tests Remaining Readings QC Pending System Messages Event Time Row Item Automatic Clot Calibration Maintenance In Tray Volume Calibrated # Tests # Available Cuvette Blank Lido System Flush Tip Cleaning Reaction Tray Full Fluid Levels Severity Level Observation Aux. Off +BUP Calibration Required System Alerts +C15 Reagent Blank Required System Log 100 % +C30 +EX3 Connection Waste (S3) Sys. Water (S4) Cleaner +EX5 Offline Operating LIMS System Water Cleaner Auxiliary Pumps 11 % Healthineers 50 % Atellica DT 250 Quality Control Reaction pending acc. Control Type O Qualitative Quantitative 6AM/XTC Level 4 Controls Level 0 Control Vial Id Level 5 y1 SpMD Level 4 Lot Number SpMD level 3 Pending Expiration Statistics Test Pos / Neg Min Separation Replicate Quick load 6AM 10Q Positive Pending Calibration AMPH 1KQ XTC 300Q AMPH 300Q Pending QC Done XTC 500Q AMPH 500Q BARB 200Q BARB 300Q Cumulative BENZ 200Q Historic BENZ 300Q BUPN 5Q COC 150Q Maintenance Required Twin QC COC 300Q LSD 0.5Q Cuvette blank Required METH 150Q METH 300Q System flush Required Scheduler MTQ 300Q OPI 2KQ Tip cleaning Required OPI 300Q OXY 100Q OXY 300Q Set PCP 25Q Delete Test New Edit Delete Copy Atellica DT 250 New Control Profile dat Cancel Min Separation Replicate Atellica DT 250 Settings Data Reports Methods User Management Change Password Import Data Backup QC Sets Shrink Database Calibrator Sets Delete Patient Units Conversion Reconnect Translator Close Program Open Program Files (x86) DataBase_LOGA_ANCHOR Atellica DT 250 Import QC Sets Status New Lot Number New Expiration Current 6AM/XTC Level Update ETOH LoWa DT 250 Update ETOH High ETOH LOw O Quantitative benz 1/ 1/2027 Mean Units Low Limit High Limit ETOH mg/dL AMPH 300S AMPH 500/1KS BARB 200/300S BENZ 200/300S BUPN 5S COC 150/300S CREATININE CSA EDDP 100Q EDDP 300Q EDDPs ETGs FENT LSD 0.5S METH 150S METH 300S MPA DAT1 Disclaimer Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc ., or its affiliates. All other trademarks are the property of their respective owners Unrestricted | Published by Siemens Healthineers AG | @ Siemens Healthineers AG, 2025 Siemens Healthineers HQ | Siemens Healthineers AG Siemensstr. 3 91301 Forchheim Germany Phone: +49 9191 18-0 siemens-healthineers.com
- atellica
- dt
- 250
- analyzer
- system
- overview
- hardware
- software
- drug
- testing