Learn how to move the Siemens Healthineers multitilt patient table to the CPR position using the Pilot control module. All steps are demonstrated on ARTIS icono biplane.

Target group: Basic user
Audio: Yes
Recommended to be viewed on the following devices: All (incl. tablet, smartphone)

Siemens Health and EARS patient tables are designed for cardio pulmonary resuscitation, CPR in short with a pressure compression of up to 60 kilograms. If you need to perform a cardiac massage, the table shouldn't be tilted. The tabletop must be moved back towards the supporting table columns. This warning label indicates the position of the tabletop for CPR. Before starting CPR, make sure that there's enough space for performing CPR. Remove any objects from the examination area. If block movements is active, press the CPR key twice and keep it pressed until you can access the patient or until the movement stop. If you don't have enough space, rotate the patient table if necessary. Don't apply horizontal forces to the table during the CPR procedure as the tabletop may move.

Move the multitilt patient table to the CPR position ARTIS icono biplane SIEMENS 40kg Healthineers 60kg System Pos Patient Transf Head Side Biplane Head Sid OK Left Side Biplane Left Side Parking / Self Test Please note this learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Unrestricted | Published by Siemens Healthineers AG | Siemens Healthineers AG, 2024 Siemens Healthineers HQ I Siemens Healthineers AG Siemensstr. 3 91301 Forchheim Germany Phone: +49 9191 18-0