ARTIS icono floor, ARTIS icono ceiling, ARTIS pheno - Remove and insert the scatter radiation grid (USA)
This job aid explains how to remove and insert the scatter radiation grid in the flat detector.
Target group: Basic user
Recommended to be viewed on the following devices: All (incl. tablet, smartphone)
Remove and insert the scatter radiation grid ARTIS icono floor, ARTIS icono ceiling, ARTIS pheno This document explains how to remove and insert the scatter radiation grid. Removing the scatter radiation grid reduces radiation dose and affects image quality. Acquisition without the grid can be appropriate, for example, for pediatric examinations. 1. Make sure that the system is in an anteroposterior (AP) position to prevent the grid from falling down. 2. A button is located on the side of the flat detector (FD) housing. Press the button. The grid slides out of its holder. - - VI -(1) (3) -(2) (1) Grid slot (2) Grid (3) Button for grid release Unrestricted | Published by Siemens Healthineers AG © Siemens Healthineers AG, 2024 SIEMENS Master Template Effective Date: 21 Apr 2023 | HILS 2224 Job Aid Effective Date: 15 May 2024 | QR700008445 Healthineers Job Aid | Remove and insert the scatter radiation grid Effective Date: 15 May 2024 | QR700008445 3. Carefully pull the grid out from the holder. 4. Put the grid in a safe place and ensure that you don’t drop or damage it. A damaged grid can reduce image quality. 5. After removing the grid, check that the icon for scatter radiation grid inserted in the Info Area on the display is struck through. 0.0 mGy 0.0 mGy E6 000.0 6. Re-insert the grid as soon as you have completed the examination to ensure optimum image quality during standard examinations. 7. Slide the grid into the detector housing until it audibly engages. 8. Ensure that the labels on the grid are facing toward the patient. Ensure that the arrows on the grid point toward the detector. 9. After inserting the grid, check that the icon for scatter radiation grid inserted is no longer struck through. 0.0 mGy 0.0 mGy E0 000.0 Unrestricted | Published by Siemens Healthineers AG © Siemens Healthineers AG, 2024 SIEMENS Master Template Effective Date: 21 Apr 2023 | HILS 2224 Job Aid Effective Date: 15 May 2024 | QR700008445 Healthineers Page 2 of 4 Job Aid | Remove and insert the scatter radiation grid Effective Date: 15 May 2024 | QR700008445 Grid types and SID range The system automatically detects the type of the inserted grid. You can use the alternative grid for cardiac and body examinations using the small FD. The alternative grid can potentially enhance image quality more than standard grids, without requiring any changes to the SID. Please note: Do not use the alternative grid for 3D acquisitions. Detector type Grid type Maximum SID range [mm] All No grid 900-1300 Small FD Standard 15/80 900-1240 Alternative grid 29/80 1200 Large FD Standard 15/80 900-1240 or 1000-1300 Unrestricted | Published by Siemens Healthineers AG © Siemens Healthineers AG, 2024 SIEMENS Master Template Effective Date: 21 Apr 2023 | HILS 2224 Job Aid Effective Date: 15 May 2024 | QR700008445 Healthineers Page 3 of 4 Job Aid | Remove and insert the scatter radiation grid Effective Date: 15 May 2024 | QR700008445 Please note this learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Siemens Healthineers HQ I Siemens Healthineers AG Siemensstr. 3 91301 Forchheim Germany Phone: +49 9191 18-0 siemens-healthineers.com Unrestricted | Published by Siemens Healthineers AG © Siemens Healthineers AG, 2024 SIEMENS Master Template Effective Date: 21 Apr 2023 | HILS 2224 Job Aid Effective Date: 15 May 2024 | QR700008445 Healthineers Page 4 of 4
- artis
- icono
- artisicono
- icon
- floor
- interventional
- IR
- neuro
- neurology
- cardiology
- card
- interventionalradiology