ADVIA® 2120i Replace the Sheath Filters Job Aid

This job aid explains how to replace the sheath filters​ on the ADVIA® 2120i System.

ADVIA 2120i Replace the Sheath Filters – Bi-Monthly Open the right and left front doors of the analyzer. There are 2 sheath filters located on either side of the analyzer. The perox sheath filter is in a housing on the left side (A) and the RBC/baso sheath filter is in a housing on the right side (B). (Continued on next page) B A ADVIA' 120 ADVIA 21201 ADVIAZIZ ADVIA 120 ADVIA 21201 PEROX 2 PEROXY RÉnceil P9221 PEROX 2-<< 38"E) ADVIA PERDE 140 M(* ADVIA 21201 PEROZ SWEATH 575 m 2325 ml. 15' 15:01 0101382 001334 99816 001360 2013-02 2013-47 2013-01 2013-02 SIEMENS Effective Date: 18-May-2022| HOOD05162003277428 Unrestricted © Siemens Healthineers, 2022 1 ADVIA 2120i Replace the Sheath Filters (Continued) 1. Carefully pull the filter toward you enough to remove the filter from the mounting and gain access to the tubing. 2. Place paper towels under the filter to catch any fluid that leaks out when you remove the filter. 3. Disconnect the reagent line attached to the barbed fitting at the input port of the filter. 4. Turn the luer fitting counter-clockwise to disconnect it from the output port. 3 Do not pull the tube out of the the luer fitting. 5. Discard the filter according to local environmental regulations. 6. Hold the replacement sheath filter by the body and attach the luer fitting to the connector. 7. Turn the luer fitting clockwise to tighten. 8. While still holding the filter by the center of the body, connect the reagent line to the barbed fitting of the input port. 9. Press the sheath filter into the mounting clip with the input end of the filter facing up. 4 10. Prime the reagent lines 5 times. 11. Check the level of the liquid in the sheath filter. 12. If the filter is not approximately 90% full, run 2 additional prime cycles. Notes: • The plastic barbed fitting and the luer connector are very fragile. Always hold the sheath filter by the body and do not use excessive force when connecting the lines. Do not use a syringe to prime the filter. A syringe can damage the filter. 2 • Effective Date: 18-May-2022| HOOD05162003277428 Unrestricted © Siemens Healthineers, 2022 SIEMENS Healthineers Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. © Siemens Healthcare GmbH 2022 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens-healthineers.com 3 Effective Date: 18-May-2022| HOOD05162003277428 Unrestricted © Siemens Healthineers, 2022

  • 120
  • 2120
  • advia2120i
  • advia2120