ADVIA® 2120i Obtain Background Counts Job Aid
This job aid explains how to obtain background counts on the ADVIA® 2120i System.
ADVIA 2120i SIEMENS Healthineers Obtain Background Counts Background Counts About Run a background count cycle to obtain a BASO WBC background count, a platelet background count, BASO WBC < = 0.10 and an HGB Trans. O 1. Navigate to Instrument. PLT < = 5 2. In the Background Counts section, select Refresh. O HGB Trans 2.5 to 4.1 The background results are color-coded. O • Green: Within range Date 12/Apr/2022 • Red: Out of range Time 08:15:13 If any result is out of range, select Refresh to run another background count cycle. If any result is still unacceptable, perform a system wash. Start 2120i Standby 2120i Startup Report 2 Refresh AutoSampler Stop/Start Eject Rack Automation Online 1 Effective Date: 18-May-2022| HOOD05162003277428 Unrestricted © Siemens Healthineers, 2022 SIEMENS Healthineers Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. © Siemens Healthcare GmbH 2022 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens-healthineers.com 2 Effective Date: 18-May-2022| HOOD05162003277428 Unrestricted © Siemens Healthineers, 2022
- 120
- 2120
- advia2120i
- advia2120