ADVIA® 2120i Hematology System Adjust the Hemoglobin Lamp Video

This video demonstrates how to adjust the hemoglobin lamp on the ADVIA® 2120i Hematology System.

Welcome to the Advia 2120I. Adjust the hemoglobin. Lamp. Video. The hemoglobin lamp assembly derives the hemoglobin. Concentration of a sample. The lamp's. Power is measured by the transmission reading. The hemoglobin. Lamp needs to. Be. Adjusted after the lamp is replaced. When the reading is greater than or equal. To 4.2. Or. When indicated during. Troubleshooting. This video will demonstrate. How to? Adjust the hemoglobin lamp. To adjust the hemoglobin lamp. Select the procedures icon and then select adjustments. Select hgb. Baseline. Select start cycle. The analyzer is now checking the baseline value to access. The colorimeter open the front doors. Of the. Analyzer. The colorimeter. Is located above the unified fluidic. Circuit. Insert a small Flathead screwdriver. Into the hole on the top right corner of the colorimeter. Use the screwdriver. To turn the adjustment screw inside. To increase the value. Turn the screw clockwise. To decrease the value. Turn the screw counter clockwise. As you turn the screw. The current reading updates. Once the current value is 4. .1. Select exit.

00 4.1 ADVIA® 2120i Hematology System Adjust the Hemoglobin Lamp Master Template HILS2218 | Effective Date: 25 Mar 2022 Adjustments Admin Operations Procedures Adjust Lamp/Temp Next Sample : Human Patient CBC / DIFF System Preparation Completed Adjustments Information Perox Lamp You can use the Adjust Lamp/Temp tab to perform adjustment procedures on the alignment of the Perox lamp, HGB Baseline, and Baso and Perox HGB Baseline temperatures. Choose the associated button and select ' Start Cycle ' to begin the procedure. When you are finished, choose Exit to exit and return to the Introduction window. Baso Temperature Perox Temperature Instrument Adjust HGB Baseline to 4.0 +/- 0.1 A/D Values Lowest: Current: Highest: Start Cycle Exit Logs SIEMENS ADVIA 21201 Disclaimer Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. liable for damages. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. ADVIA, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc ., or its affiliates. All other trademarks and brands are the property of their respective owners. Siemens Healthcare GmbH 2023 91052 Erlangen, GermanylTelephone: +49 9131 84-Olsiemens-healthineers.com

  • adiva
  • avdia
  • advai
  • 2021