Siemens Healthineers Academy

ADVIA 2120i Hardware Overview Video

This video describes the hardware components of the ADVIA 2120/2120i Hematology Analyzer.

Welcome to the ad via 2120 I. Hardware overview. This video will review the main external and internal components of the analyzer. There are four main covers that you will use to access internal parts of the analyzer. The right front door. The left front door. The bottom door. And the top lid. The touch pad controls the main functions of the analyzer. It is located on the right side of the analyzer, just below the right front door. There are five buttons. On stand by. Start or stop the autosampler. Eject the rack. And power off. Each button also has an indicator light except for the power off button. There is also an indicator light that lights up when a rack is in the auto sampler. You can use the barcode reader to scan the barcodes on reagents. Sample tubes control information sheets. And more. To use the barcode reader pointed at the barcode. As the barcode is read, the reader beats and the LED light flashes. This is the autosampler input tray. When you place racks in this tray, the instrument will automatically pull them into the autosampler and begin processing samples. When all of the samples on a rack have been aspirated, it will be ejected. Here on the output tray. The manual closed tube sampler is located on the front of the analyzer. There is a needle inside the centering collar. When you press a tube onto the needle, it punctures the tube closure and aspirates the sample. This process is described in greater detail in other learning objects. The manual open tube sampler is located on the front of the analyzer. To use it, immerse the sample probe in the specimen and press the aspiration plate. This process is described in greater detail in other learning objects. This is the unified fluidic circuit, also referred to as the UFC. It is located in the front center of the analyzer and is visible through the left front door. The unified fluidic circuit is an acrylic block that contains pathways, valves, and reaction chambers. It simplifies maintenance by eliminating the need for multiple lengths of tubing connections and valves. The hemoglobin colorimeter measures the light that passes through the reaction chamber. The analyzer uses the readings to derive the hemoglobin concentration. This is the sheer valve. After a sample is aspirated, it is drawn into the sheer valve. The valve rotates to divide the sample into aliquots for different types of tests. Reagent bottles are located behind the right front door and left front doors. Reagents are either provided in individual bottles or in bundles of bottles. A bundle of connected bottles is called a time pack. The diff time pack is composed of the perox sheath and the Perox 1/2 and three reagents. It is located on the left side. The CBC Time pack is composed of the RBC PLT, Baso and AGB reagents. It is located on the right side. There is also a cyanide free version of this time pack. In that version, the HBB reagent is replaced with a cyanide free version. The auto retake and easy wash bottles are located to the right of the CBC time pack. The Deformer bottle is located above the easy wash bottle. The overflow bottle is located next to the Deformer bottle. If a reaction chamber is blocked, the overflow liquid will flow into the overflow bottle. During normal operation, you may occasionally see some liquid in the overflow bottle. However, if the bottle frequently fills with liquid contact Siemens healthineers support. There are four pumps, also referred to as syringes. They are located on either side of the unified fluidic circuit. The syringes move up and down to move fluids throughout the analyzer. The two outer syringes are the sheath rinse pumps. The inner syringe on the left is the parox sample pump. The inner syringe on the right is the RBC Baso and Retic sample pump. The RBC Optical Assembly is located on the right side of the analyzer under the top lid. It uses a laser and detector to measure the red blood cells reticulocytes and platelets that pass through the assembly. The Perox Optical Assembly is located on the left side of the analyzer under the top lid. It uses a lamp and detector to measure the white blood cells that pass through the assembly. The sheath rinse is located on the floor or on the bench below the analyzer. The sheath rinse is fluid that envelops the sample stream as it passes through the different components of the analyzer. It is also used to clean the hydraulic pathways and reaction chambers after each sample to prevent carryover and ensure the integrity of the results. Finally, the waste bottle is located below the analyzer. Your analyzer can have a manual waste bottle or an auto waste system. The auto waste system automatically empties the waste bottle as it fills. A manual waste bottle must be emptied manually according to your laboratory procedures. The process to manually empty the waste is described in greater detail in another learning object.

100 120 20 053884 053882 053820 2013-07 19227 2023-02-28 2023-12-3 8010 80101 ADVIA® 2120i Hardware Overview SIEMENS Opticon ADVIA ADVIA 120 ADVIA 21201 PEROX 1 PEROX 2 PEROX3 575 m 585 mL 650 m EZ WASH 注意 REV. A ASSY 20L/5 GAL Disclaimer Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Siemens Healthcare GmbH 2022 91052 Erlangen, GermanylTelephone: +49 9131 84-Olsiemens-healthineers.com

  • 120
  • 2120
  • advia2120i
  • advia2120