ADVIA® 2120i Analyze Samples Job Aid

This job aid explains how to analyze samples on the ADVIA® 2120i System.

ADVIA 2120i SIEMENS Healthineers Analyze Samples Analyze samples with the Autosampler 1. Insert the tube into the rack with the barcode label visible above the rack barcode label that indicates the rack number and sample position. • Do not twist the tube within the rack. If the tube needs to be repositioned, remove it from the rack first, reposition the tube, then reinsert the tube into the rack. 2. Load the rack onto input queue with the labels facing the front of the analyzer. (Continued on next page) Note: Load samples in the following order: 1. Optional: Whole blood primer (with whole blood primer label) 2. Controls 3. Patient samples (with sample ID labels) Note: Workorders are created within the Laboratory Information System 1 (LIS). The LIS is queried when a sample ID is read by the analyzer. Effective Date: 18-May-2022| HOOD05162003277428 Unrestricted © Siemens Healthineers, 2022 ADVIA 2120i SIEMENS Healthineers Analyze Samples (continued) 3. If the Standby indicator on the touchpad is lit, this indicates the analyzer is in standby mode. Press the Standby button to resume normal operation. 4. On the touchpad, press Start/Stop Sampler (Continued on next page) 3 4 001 2 Effective Date: 18-May-2022| HOOD05162003277428 Unrestricted © Siemens Healthineers, 2022 ADVIA 2120i SIEMENS Healthineers Analyze Samples (continued) Analyze a whole blood sample with the Manual Closed - Tube Sampler If the Standby indicator on the touchpad is lit, this 3 Specimen Type O Whole Blood O Body Fluid O CSF 1. indicates the analyzer is in standby mode. Press the Standby button to resume normal operation. 2. On the workstation PC, navigate to Operations > 4 Sample ID ABC1234 Cancel Manual Sample . 3. Select Whole Blood in the Specimen Type box. Auto Increment OK 4. Use the barcode reader or keyboard to enter the Note: Sample ID . • If you manually enter a Sample ID, you must 5 Sample Type O Patient O Control O Primer Run samples in the following order: click the OK button. • Controls 5. Select a Sample Type • Patient samples (with . 6. Select a Selectivity 6 Selectivity O CBC O CBC/DIFF O CBC/DIFF/RETIC O CBC/RETIC O RETIC sample ID labels) . 7. Select a Species Type Optional: Whole blood . • • You can only select a Species Type on a 7 Species Multispecies analyzer primer (with whole (Continued on next page) Species Types 0 Human M Sort By Usage blood primer label) Note: Run samples in the following order: Body Fluid Rinse 1. Controls 2. Patient samples (with sample ID labels) 3. Optional: Whole blood primer (with 3 whole blood primer label) Effective Date: 18 -May -2022| HOOD05162003277428 Unrestricted © Siemens Healthineers, 2022 ADVIA 2120i SIEMENS Healthineers Analyze Samples (continued) 8. Mix the sample. 9. Turn the tube upside down and press it down into 9 the sampler. • Hold the tube straight and parallel to the centering collar wall. 10. The sample is automatically aspirated, and the sampling light flashes. 10 11. When the sampling light stops flashing, remove the tube. 02 (Continued on next page) 130 4 Effective Date: 18-May-2022| HOOD05162003277428 Unrestricted © Siemens Healthineers, 2022 ADVIA 2120i SIEMENS Healthineers Analyze Samples (continued) Analyze a whole blood sample with the Manual Open - Tube Sampler 1. If the Standby indicator on the touchpad is lit, this 3 Specimen Type O Whole Blood O Body Fluid O CSF indicates the analyzer is in standby mode. Press the Standby button to resume normal operation. 2. On the workstation PC, navigate to Operations > 4 Sample ID ABC1234 Cancel Manual Sample . 3. Select Whole Blood in the Specimen Type box. Auto Increment OK 4. Use the barcode reader or keyboard to enter the Note: Sample ID If you manually enter a Sample ID, you must 5 Run samples in the . Sample Type O Patient O Control O Primer • following order: click the OK button. • Controls 5. Select a Sample Type Select a Selectivity Selectivity • Patient samples (with . 6. 6 O CBC O CBC/DIFF O CBC/DIFF/RETIC O CBC/RETIC O RETIC . 7. Select a Species Type sample ID labels) Optional: Whole blood . • • You can only select a Species Type on a 7 Species Multispecies analyzer primer (with whole . Species Types 0 Human M Sort By Usage blood primer label) (Continued on next page) Note: Body Fluid Rinse Run samples in the following order: 1. Controls 2. Patient samples (with sample ID labels) 5 Effective Date: 18 -May -2022| HOOD05162003277428 Unrestricted © Siemens Healthineers, 2022 ADVIA 2120i SIEMENS Healthineers Analyze Samples (continued) 8. Ensure the sample is well mixed. 9. Immerse the sampler probe only deep enough to ensure aspiration. 11 10. Press the aspirate plate. 11. The sampling light flashes during aspiration. 12. When the sampling light stops flashing, remove the sample. 10 6 Effective Date: 18-May-2022| HOOD05162003277428 Unrestricted © Siemens Healthineers, 2022 SIEMENS Healthineers Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. © Siemens Healthcare GmbH 2022 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens-healthineers.com 7 Effective Date: 18-May-2022| HOOD05162003277428 Unrestricted © Siemens Healthineers, 2022

  • 120
  • 2120
  • advia2120i
  • advia2120